Skip to main content
Menu
US
Cell Counting and Image Cytometry

Antibody Neutralization

Section
Celigo Applications
Celigo Applications
Cell Counting Method Selection
Cell Counting and Image Cytometry FAQs
Cell-based Assays for Bioprocessing
Cell-based Assays for Gene Therapy Development
Cellometer Applications
Modern Virology Assays
Sub Section
Assays for Virology Research
3D Cell Models
Assays for Virology Research
Brightfield Applications
Celigo Cell Counting Applications
Celigo Fluorescent Assays
Celigo Image Cytometer Direct Cell Counting Assays for Immunotherapy
Cell Line Development
Migration and Invasion Assays
iPSC Reprogramming
Topic
Antibody Neutralization
Antibody Neutralization
CAR T Cell-Mediated Cytotoxicity
Viral Infection
Measure Viral Titers
Plaques and Foci Counting
Transduction Efficiency

Measure antibody neutralization using direct cell counting in 96-well microplates using the Celigo image cytometer

  1. High-throughput automated imaging and analysis directly in plates
  2. Directly count infected cells in plates without trypsinization
  3. Measure antibody neutralization effect in 8 min

Introduction to measuring antibody neutralization

Antibody neutralization assays utilize antibodies or directly with serum to prevent infectivity of viruses. In general, the host cells are treated with different titrations of the neutralizing antibodies or sera. Next, the viruses are allowed to infect the host cells at different infectivity. Finally, each titration is analyzed to determine the infectivity and quantify the neutralization capabilities of the antibodies or sera.

Protocol

  1. First the virus particles (RVPs) are incubated with different titrations of human sera for 1 hour at 37 °C
  2. Raji cells are then infected with the antibody-RVP complexes and seeded in a 96-well microplate
  3. The plate is then incubated at 37 °C for 48 hours
  4. After incubation, the plate is imaged and analyzed (8 min/96-well plate) using the Celigo™ image cytometer to count the number of GFP positive infected cells
  5. The results are plotted as a dose response of sera titration and the IC50 values are calculated

Measure antibody neutralization of different antibodies on Zika virus using direct cell counting in 96-well plates

  1. Raji cells and 6 different antibody-RVP complexes (Zika virus) are infected and seeded in 96-well microplates following the protocol described above
  2. After 48 hours, the Celigo image cytometer was used to count the number of GFP positive infected cells in each well to generate a dose response curve for all tested antibody-RVP complexes
  3. Two sets of example images of antibody-RVP complexes treatment are shown below, where GFP positive infected cells can be observed at different antibody concentrations

Brightfield and fluorescent overlay images of Zika virus-infected cells

Antibody neutralization-1

Example images of antibody-RVP complexes treatment showing an increase
in the number of infected (GFP positive) cells as
antibody concentrations decreased.

Directly count cells in images

Antibody neutralization-2

The Celigo was used to count the total number of cells in brightfield and identify the GFP positive cells in histogram gating.

1. After counting the number of infected cells and total cell count for each antibody-RVP complex and concentrations using the gating function in the Celigo, a dose-response plot of percent infection is generated and IC50 values are calculated.

Green: Mean Intensity Histogram

Antibody neutralization-3

2. The dose-dependent antibody neutralization IC50 results are directly compared to FACS and are highly comparable.

Celigo

Antibody neutralization-4

Antibody neutralization dose-response curves for 6 antibodies using the Celigo.

FACS

Antibody neutralization-5

Antibody neutralization dose-response curves for 6 antibodies using FACS.

IC50 (µg/ml) Ab1 Ab2 Ab3 Ab4 Ab5 Ab5
FACS 0.0077 0.0089 0.0017 0.011  0.547 4.683
Celigo 0.0090 0.0081 0.0042 0.011 0.489 0.307

Measure antibody neutralization of different sera on cytomegalovirus using direct cell counting in 96-well plates

  • The ARPE epithelial host cells are seeded in 96-well microplate and incubated for 24 hours to allow the cells to adhere
  • The cells are then infected by cytomegalovirus and incubated with titrations of different sera for another 24 hours
  • Finally, the cells are stained with primary antibodies conjugated with Alexa Fluor 647 (AF647) or Horseradish Peroxidase (HRP)
  • After staining, the cells are imaged and analyzed using the Celigo image cytometer to count the number of infected cells

Brightfield-based antibody neutralization assay using horseradish peroxidase (HRP)

Brightfield images of HRP-stained infected cells at high serum

Antibody neutralization-6

The example image shows the low number of infected cells with high
serum concentration in brightfield using HRP

Brightfield images of HRP-stained infected cells at low serum

Antibody neutralization-7

The example image shows the high number of infected cells with low
serum concentration in brightfield using HRP

Whole plate image and antibody neutralization dose-response curves of 4 sera

Antibody neutralization-8

The number of infected cells is counted and dose response curves are generated for 4 different sera, where the IC50 values are calculated. The Celigo is able to output the counted results directly to Microsoft Excel for analysis.

Fluorescence-based antibody neutralization assay using AF647

Fluorescent images of AF647-stained infected cells at high serum

Antibody neutralization-9

The image shows the low number of infected cells with high serum 
concentration in fluorescence using AF647

Fluorescent images of AF647-stained infected cells at low serum

Antibody neutralization-10

The image shows the high number of infected cells with low serum concentration 
in fluorescence using AF647

Whole plate image and antibody neutralization dose-response curves of 4 sera

Antibody neutralization-11

The number of infected cells is counted and dose response curves are generated for 4 different sera, where the IC50 values are calculated. The Celigo is able to output the counted results directly to Microsoft Excel for analysis