Skip to main content
Menu
US
Cell Counting and Image Cytometry

Cell Count and Viability Measurement for Clinical Primary Samples in Immunology

Section
Cellometer Applications
Celigo Applications
Cell Counting Method Selection
Cell Counting and Image Cytometry FAQs
Cell-based Assays for Bioprocessing
Cell-based Assays for Gene Therapy Development
Cellometer Applications
Modern Virology Assays
Sub Section
Immuno-oncology Research Using Cellometer Instruments
Automated Analysis of Primary Cells
Brightfield Cell Analysis
Cell Analysis for Breweries and Biofuels
Cell Concentration and Viability for Blood-based Samples
Cell Counting with a Hemocytometer
Cell Viability for Clean and Messy Samples
Fluorescent Assays
Immuno-oncology Research Using Cellometer Instruments
Miscellaneous Cell Types
Topic
Cell Count and Viability Measurement for Clinical Primary Samples in Immunology
Adoptive Cell Transfer Therapy Research
Cell Count and Viability Measurement for Clinical Primary Samples in Immunology

Introduction to immunology

The field of immunology encompasses many different areas of research and therefore has a very broad range of primary samples - from human primary and clinical samples to immobilized cell lines to samples from murine models.

Although there is a great variety of clinical and primary samples, most often the concentration and viability is determined by a single method: hemocytometer counting using Trypan blue. This counting method may not be optimal for all samples. Samples such as bone marrow, cord blood, tissue digests, bronchoalveolar lavage, may be contaminated with residual red blood cells, tissue and cellular debris. These contaminants can lead to inaccurate enumeration of immune cells.

Revvity's solution is to perform fluorescent based cell analysis using acridine orange (AO) and propidium iodide. Since AO and PI are nucleic acid dyes, they will bind to and label only the DNA of live and dead nucleated cells. This method allows for the identification and, enumeration, as well as measuring viability of nucleated cells in any sample.

Strategy for the detection of nucleated cells in complex primary samples

Directly measure total nucleated cells using acridine orange (AO) and propidium iodide (PI) staining and Cellometer automatic cell counter

  1. Stain sample with AO/PI at 1:1 ratio
  2. Load 20 µl* stained cell sample into counting chamber (*10 µl for Ascend 8-chamber)
  3. Count and analyze sample using a Cellometer automated fluorescent viability counter (Ascend, Auto 2000, K2, Spectrum)

Measuring cell concentration and viability of primary cell samples

Quantifying cell samples in 4 easy steps using fluorescent viability counter

Using a Cellometer with dual fluorescence – Ascend, Auto 2000, K2 or Spectrum –a few microliters of stained sample is added to the Cellometer Counting Chamber. Imaging and analysis of the samples is completed in less than 60 seconds. Brightfield and fluorescent cell images can be viewed to check cell morphology and verify cell counting. Total cell count, concentration, and mean diameter are automatically displayed.

Cellometer slide

1. Pipette 20 µl* of stained sample into a disposable slide (* 10 µl in Ascend 8-chamber slide)

Insert slide vision

2. Insert slide into the instrument

aopi select assay

3. Select assay from a drop down menu

Click count, acquire image and view cell count, concentration, diameter

4. Click count, acquire image and view cell count, concentration, diameter


Detection of immunological cell samples using Cellometer Auto 2000

Simple, user-friendly Cellometer procedure

Using Cellometer Mini, Auto 1000, Auto T4, Auto 2000, K2 and Vision

1. Pipette 20µl 2. Insert slide 3. Select assay & click count 4. Results in 30 seconds

Human clinical research samples

Bone marrow (BM)

Bone marrow (BM) is a complex tissue that is the site of hematopoiesis and B-cell development. Multipotent, non-hematopoietic stem cells, such as mesenchymal stem cells, may be isolated from human bone marrow. Furthermore, bone marrow mononuclear cells (MNCs) can be isolated by diluting the bone marrow sample and performing a density gradient separation. After centrifugation, the MNC layer is collected and may be further processed to isolate various subpopulations (hematopoietic progenitor cells, mesenchymal stem cells, CD33+ myeloid cells, and CD138+ plasma cells) using cell sorting.

Current issue: Sample contamination with red blood cells may lead to inaccurate enumeration of immune cells.

Current solution: Lyse red blood cells prior to cell count.

Revvity solution: Perform fluorescent based analysis. Using AO/PI allows for the identification and enumeration of mononuclear cells, as well as measuring cell viability in the counted sample.

Bone marrow samples can be processed and used for many different experimental applications such as studying hematopoietic cell differentiation pathways and acting as controls for BM malignancies studies.

Total Nucleated Cells in Bone Marrow Measured Using AO

Total nucleated cells in bone marrow measured using AO

Isolated Mononuclear Cells Stained with AOPI

Isolated mononuclear cells stained with AOPI

Bone Marrow Stem Cells Stained with AOPI

Bone marrow stem cells stained with AOPI


In a separate experiment, three individual human bone marrow samples were stained with AOPI, imaged and analyzed. Acquired brightfield (BR) images were counted separately from the acquired AOPI (F1F2) images. As shown in the table (right), many more particles were counted in the BR than in F1+F2. Since the count of F1F2 to BR are not 1:1, the ratio of FL /BR reflect those counts. Furthermore, the tabulated ratios in the three separate experiments are shown to be different from experiment to experiment. The ratio variation is the direct results of imaging of non-nucleated particles in brightfield. Due to the complex nature of the sample, we conclude that counting nucleated cells in a bone marrow sample should be performed using AOPI staining.
 

  F1F2 BR Ratio FL / BR Viability
MNC-BM_1 3296 4575 0.72 98.20%
MNC-BM_2 1388 1605 0.86 98.60%
MNC-BM_3 2008 2935 0.68 98.20%

Cord blood (CB)

Cord blood (CB) is collected at the time of delivery in a 250 mL standard blood bag containing anti-coagulating agents. CB-Mononuclear Cells (MNC) can be isolated by diluting the cord blood sample and performing a density gradient separation. After centrifugation, the MNC layer is collected and can be further processed to isolate subpopulations (hematopoietic progenitor stem cells, CD34+ neutrophils, CD14+ monocyte, CD19+ B cells, CD3+, CD4+, CD8+ T cells, and CD56+ NK cells) using cell sorting.

Total Live and Dead Cells in Cord Blood Measured Using AOPI

Total live and dead cells in cord blood measured using AOPI

Isolated Mononuclear Cells Stained with AOPI

Isolated mononuclear cells stained with AOPI

Bone Marrow Stems Cells Stained with AOPI

Bone marrow stems cells stained with AOPI


Five human cord blood samples were stained with AOPI and imaged. Acquired brightfield (BR) images were counted separately from the acquired AOPI (F1F2) images. As shown in the table (right), many more particles were counted in the BR than in F1+F2. The ratios of FL to BR show that in almost every sample there was more than double the amount of particles counted in BR compared to AOPI. Variability in ratios stems from counting non-nucleated particles in brightfield. We conclude that counting nucleated cells in complex samples such as cord blood should be performed using AOPI.
 

  F1F2 BR Ratio FL / BR Viability
MNC-CB_1 2488 4659 0.53 98.10%
MNC-CB_2 2613 7152 0.37 94.10%
MNC-CB_3 2313 5791 0.40 93.30%
MNC-CB_4 1939 3449 0.56 96.70%
MNC-CB_5 2243 6450 0.35 95.80%

Human peripheral blood (aka whole blood)

Human peripheral blood (aka whole blood) is collected from a patient into a vial containing anti-coagulating agents. Whole blood is a rich source of diverse mononuclear cells. Subpopulations of nucleated cells (progenitor cells, B cells, T cells, dendritic cells, natural killer cells, neutrophils, red blood cells, platelets, monocytes, macrophages) can be isolated and purified using various isolation techniques.

Total Number Nucleated Cells in Whole Blood Measured Using AO

Total number nucleated cells in whole blood measured using AO

Isolated Mononuclear Cells Stained with AO

Isolated mononuclear cells stained with AO

Purified CD19+ B Cells Stained with AOPI

Purified CD19+ B cells stained with AOPI

Isolated Neutrophils Cells Stained with AOPI

Isolated neutrophils cells stained with AOPI

NK Cells CD56+ Stained with AOPI

NK cells CD56+ stained with AOPI

HPC (Human Hematopoietic Cells)

HPC (Human hematopoietic cells) Human hematopoietic primary cells are isolated from peripheral blood, bone marrow or cord blood.

CD34+ Stem Cells Stained with AOPI

CD34+ stem cells stained with AOPI

Murine samples

Murine animal models are widely utilized to study different human diseases. A great variety of samples are regularly collected and processed. Obtaining a cell count and viability are two common processes that are part of standard laboratory practices.

Current issue: During standard harvests, the processing of numerous samples is very time consuming. Furthermore, due to the great variety of potential samples multiple methods may be needed to determine cell count and viability.

Current solution: Hemocytometer counting.

Revvity solution: Perform fluorescent based analysis. Using AO/PI allows for the identification and enumeration, and measures cell viability of nucleated cells in the collected sample. This procedure may be performed on many different collected samples.

A large immunological study was performed where 30 individual samples of mouse lymphocytes and splenocytes were collected and processed on the same day. Each sample was stained with AO/PI and imaged using a Cellometer Auto 2000. The total cell count, concentration and viability were acquired for all of the samples.
 

  Lym7 Lmy6 Lym2 Lym5
Total cell count 49 1419 10635 24245
Live cell concentration 1.07E+05 2.84E+06 3.11E+07 6.75E+07
Viability 63.20% 58.00% 84.40% 80.30%
Brightfield cell image
Bright field cell image - Lym7
Bright field cell image - Lym6
Bright field cell image - Lym2
Bright field cell image - Lym5
Counted live / dead cell image
Counted live-dead cell image - Lym7
Counted live-dead cell image - Lym6
Counted live-dead cell image - Lym2
Counted live-dead cell image - Lym5


Above, is the representative set of 4 different mouse lymphocyte samples (Lym7, 6, 2, 5). A brightfield and a counted fluorescent image are shown for each sample. The samples ranged in concentration from 1.07x105 (with 49 cells counted) to 6.75×107 (with 24245 cells counted) and viability of samples ranged from 58.0% to 84.4%. Because images can be saved during data acquisition, they can be used for documentation and visual verification of counted cells.

Below, are the tables containing the data for each of the 30 collected and processed lymphocyte and splenocyte samples. Reported below are the cell counts, concentrations, and viability for live cells. The same imaging and counting parameters were used for all samples. Total imaging time for all 60 samples was approximately 30 minutes compared to total counting time using a hemocytometer was approximately 2 hours.
 

Sample ID Cell count Concentration Viability
Ln 1 5627 1.94E+07 83.60%
Ln 2 9283 3.20E+07 84.30%
Ln 3 11694 4.03E+07 83.20%
Ln 4 8320 2.87E+07 82.90%
Ln 5 21310 7.35E+07 81.60%
Ln 6 608 2.09E+06 49.90%
Ln 7 19 6.55E+04 52.70%
Ln 8 5944 2.05E+07 81.50%
Ln 9 1383 4.78E+06 71.60%
Ln 10 3968 1.37E+07 75.20%
Ln 11 1641 5.66E+06 69.90%
Ln 12 12408 4.28E+07 84.90%
Ln 13 1456 5.02E+06 76.40%
Ln 14 2168 7.47E+06 76.30%
Ln 15 3912 1.35E+07 80.30%
Ln 16 3876 1.34E+07 80.00%
Ln 17 3836 1.32E+07 79.80%
Ln 18 2008 6.92E+06 70.60%
Ln 19 1940 6.68E+06 68.50%
Ln 20 1535 5.29E+06 66.90%
Ln 21 2103 7.25E+06 72.00%
Ln 22 9048 3.12E+07 80.30%
Ln 23 2607 8.99E+06 71.80%
Ln 24 8240 2.84E+07 79.10%
Ln 25 3571 1.23E+07 75.50%
Ln 26 10520 3.62E+07 81.80%
Ln 27 6069 2.09E+07 82.40%
Ln 28 5913 2.04E+07 74.50%
Ln 29 6457 2.22E+07 74.00%
Ln 30 6296 2.17E+07 78.40%

 

Sample ID Cell count Concentration Viability
Sp 1 3840 1.32E+07 58.40%
Sp 2 6322 2.18E+07 64.80%
Sp 3 4055 1.40E+07 63.60%
Sp 4 4990 1.72E+07 58.70%
Sp 5 8204 2.82E+07 62.80%
Sp 6 7704 2.65E+07 60.20%
Sp 7 8209 2.83E+07 79.90%
Sp 8 8999 3.10E+07 59.90%
Sp 9 7870 2.71E+07 69.30%
Sp 10 7894 2.72E+07 69.30%
Sp 11 3967 1.37E+07 63.80%
Sp 12 2659 9.19E+06 72.30%
Sp 13 1329 4.58E+06 61.40%
Sp 14 1290 4.45E+06 60.60%
Sp 15 2215 7.64E+06 60.70%
Sp 16 2222 7.66E+06 60.60%
Sp 17 5199 1.79E+07 70.50%
Sp 18 1895 6.53E+06 74.40%
Sp 19 1769 6.09E+06 60.50%
Sp 20 3532 1.22E+07 59.50%
Sp 21 1972 6.79E+06 67.70%
Sp 22 1764 6.08E+06 62.40%
Sp 23 2020 6.96E+06 64.60%
Sp 24 1156 3.98E+06 64.90%
Sp 25 1244 4.28E+06 60.00%
Sp 26 3884 1.34E+07 76.70%
Sp 27 1728 5.96E+06 69.80%
Sp 28 2136 7.36E+06 65.10%
Sp 29 2112 7.29E+06 67.90%
Sp 30 1080 3.72E+06 65.70%

 

Mouse Tail Vein Blood

Mouse tail vein blood

Mouse Lymph Node

Mouse lymph node

Mouse Dendritic Cells

Mouse dendritic cells

Mouse Digested Ear

Mouse digested ear

Mouse Splenocytes

Mouse splenocytes

Tumor digest

Digest samples are collected from the patient at the primary source of the tumor. Solid tumors first need to be digested and processed before they can be utilized in experiments.

Current issue: Tissue digest are inherently messy samples. They may contain tissue fragments and cellular debris making it difficult to identify and count cells of interest.

Current solution: Spin-out debris and wash sample multiple times. Isolate cells of interest by magnetic bead separation technique.

Revvity solution: Perform fluorescent based analysis. AO/PI staining allows for the identification, enumeration, and provides cell viability of nucleated cells in the collected sample.

Examples of tumor digests


Human melanoma digest sample stained with AOPI

Human Melanoma Digest Sample Stained with AOPI

Bright field image, AO/PI image & bright field and AO/PI overlay image



Murine tumor digest samples stained with AO/PI

Murine Tumor Digest Samples Stained with AO/PI
Murine Tumor Digest Samples Stained with AO/PI

Brightfield and AO/PI overlay images (far right) show the labeling of nucleated cells only.


Tumor cells with lymphocytes

Tumor Cells with Lymphocytes histogram

Tumor cells with lymphocytes histogram

Tumor Cells with Lymphocytes

Brightfield and fluorescent images of lymphocyte and tumor cell mixture stained with AOPI


Lymphocytes were added to a suspension of tumor cells and stained with AO/PI. The sample data was acquired, and images automatically analyzed. The cell diameter histogram shows that there are two distinct populations in our sample: one (in red) is the smaller ~6.0 microns in diameter and the other (in green) are the tumor cells averaging about 13 micron in diameter.

Bronchoalveolar lavage (BAL)

Bronchoalveolar lavage (BAL) is a procedure in which a bronchoscope is passed through the mouth or nose into the lungs and fluid is squirted into a small part of the lung and then recollected for examination. BAL sample may contain lung and bronchial epithelial cells, macrophage, lymphocytes, and other immune cells. BAL research samples from humans or mouse may be analyzed on the Cellometer.

Current issue: BAL samples contain cellular debris make it difficult to identify and count cells of interest.

Current solution: Upon isolation and purification of immune cells, flow cytometry or manual counting is used to count the cells.

Revvity solution: Perform fluorescent based analysis. Using AO/PI allows for the identification and enumeration, and measures cell viability of nucleated cells in the collected sample.

Examples of bronchoalveolar lavage samples

Examples of Bronchoalveolar Lavage Samples
Examples of Bronchoalveolar Lavage Samples
Examples of Bronchoalveolar Lavage Samples

These murine BAL samples were collected from multiple individual experiments. Each sample was stained with AO and imaged. Above is a collection of brightfield, AO fluorescent (FL), and BR-FL overlay images.

Counting PBMC samples

  • Analysis of peripheral blood mononuclear cells is similar to analysis of nucleated cells from bone marrow, cord blood, or whole blood
  • Current PBMC enumeration method using hemacytometer and light microscopy can include donor-dependent RBC-induced counting error
  • Cellometer image cytometer was used to eliminate the RBC-induced counting error by automatically enumerate AO/PI-stained nucleated cells

Conclusions and Cellometer selection guide

Primary samples are routinely used for biology, immunology, and cancer research. The diversity and the quantity of the samples often poses a challenge of counting and calculating the concentrations and viability of samples in a time efficient and accurate manner. The family of Cellometer instruments provides researchers multiple tools to perform fast and consistent counts of nucleated cells. Fluorescent staining of blood-based and primary samples allows for more accurate identification and enumeration of nucleated cells even in the presence of potential contaminants. The chart below provides a quick reference guide to which Cellometer may be right for your lab!
 

  Cellometer Auto T4 Cellometer Auto 2000 Cellometer Spectrum
Purified cells– no debris ++ +++ +++
Isolated MNCs – without lysing RBCs - +++ +++
Fresh BM, CB, WB, MNCs w/ lysing of RBCs + +++ +++
Fresh BM, CB, WB without lysing _ +++ +++
Frozen BM, CB, WB + +++ +++
Digested tumors and BAL with tissue/cellular debris   +++ +++

+++ = excellent, ++ = very good, + = good, - = not recommended, BM = Bone Marrow, CB = Cord Blood, WB = Whole Blood, MNCs = Mononuclear Cells

Publications using Cellometer for immunology research

  • Singh H, Figliola MJ, Dawson MJ, et al. Reprogramming CD19-specific T cells with IL-21 signaling can improve adoptive immunotherapy of B-lineage malignancies. Cancer Research 2011, 15(10):3516-27
  • Chan LL, Zhong XM, Pirani A, Lin B. A novel method for kinetic measurements of rare cell proliferation using Cellometer image-based cytometry. Journal of Immunological Methods 2012, 377(1-2):8-14
  • Holmuhamedov EL, Czerny C, Beeson CC, Lemasters JJ. Ethanol Suppresses Ureagenesis in Rat Hepatocytes: role of acetaldehyde. Journal of Biological Chemistry 2012, 287(10):7692-7700
  • Ranguelova K, Rice AB, Khajo A, Triquigneaux M, Garantziotis S, Magliozzo RS, Mason RP. Formation of reactive sulfite-derived free radicals by the activation of human neutrophils: An ESR study. Free Radical Biology & Medicine 2012, 52:1264-1271
  • Lugli E, Gattinoni L, Roberto A, Mavilio D, Price DA, Restifo NP, Roederer M. Identification, isolation and in vitro expansion of human and nonhuman primate T stem cell memory cells. Nature Protocols 2012, 8(1):33-42
  • Lo Surdo J, Bauer SR. Quantitative Approaches to Detect Donor and Passage Differences in Adipogenic Potential and Clonogenicity in Human Bone-Marrow-Derived Mesenchymal Stem Cells. Tissue Engineering Part C Methods 2012, 18(11):877-89
  • Chan LLY, Laverty DJ, Smith T, Nejad P, Hei H, Gandhi R, Kuksin D, Qiu J. Accurate measurement of peripheral blood mononuclear cell concentration using image cytometry to eliminate RBC-induced counting error. Journal of Immunological Methods 2013, 388(1-2):25-32
  • Yang B, Pham TH, et al. Accurate and Simple Measurement of the Pro-inflammatory Cytokine IL-1β using a Whole Blood Stimulation Assay. Journal of Visual Experimentation 2011 1(49)


For research use only. Not for use in diagnostic procedures.

ON THIS PAGE